Untitled Document
 
Untitled Document
 
Technical Specifications for Non-Autoclavable Eight Channel Pipette:

Catalog No.
Volume
Range
Dispensing
Increment

Test
Volume

Maximum
permissible
systematic
error
(inaccuracy)
Maximum
permissible
random error
(imprecision)
%
µl
%
µl
1010801000
0.5 - 10µl
0.1µl
10µl
1.50
0.15
1.50
0.15
5µl
2.50
0.125
2.50
0.125
1µl
4.00
0.04
4.00
0.04
1010802000
5 - 50µl
0.5µl
50µl
1.00
0.50
0.50
0.25
25µl
1.50
0.375
1.00
0.25
5µl
3.00
0.15
2.00
0.10
1010803000
50 - 300µl
5µl
300µl
0.70
2.10
0.25
0.75
150µl
1.00
1.50
0.50
0.75
50µl
1.50
0.75
0.80
0.40

Technical Specifications for Non-Autoclavable Twelve Channel Pipette:

Catalog No.
Volume
Range
Dispensing
Increment

Test
Volume

Maximum
permissible
systematic
error
(inaccuracy)
Maximum
permissible
random error
(imprecision)
%
µl
%
µl
1011201000
0.5 - 10µl
0.1µl
10µl
1.50
0.15
1.50
0.15
5µl
2.50
0.125
2.50
0.125
1µl
4.00
0.04
4.00
0.04
1011202000
5 - 50µl
0.5µl
50µl
1.00
0.50
0.50
0.25
25µl
1.50
0.375
1.00
0.25
5µl
3.00
0.15
2.00
0.10
1011203000
50 - 300µl
5µl
300µl
0.70
2.10
0.25
0.75
150µl
1.00
1.50
0.50
0.75
50µl
1.50
0.75
0.80
0.40

 


                   
                   
Untitled Document

Home    About us    FAQ    RMA    Warranty    Feedback    Contact Us    Disclaimer    Terms of Use    Privacy Policy
Copyright © 2009-2011 Catalyst Biotech. All rights reserved.