Technical Specifications for Autoclavable Eight Channel Pipette:

Catalog No.
Volume
Range
Dispensing
Increment

Test
Volume

Maximum
permissible
systematic
error
(inaccuracy)
Maximum
permissible
random error
(imprecision)
%
µl
%
µl
1020801000
0.5 - 10µl
0.1µl
10µl
1.50
0.15
1.50
0.15
5µl
2.50
0.125
2.50
0.125
1µl
4.00
0.04
4.00
0.04
1020802000
5 - 50µl
0.5µl
50µl
1.00
0.50
0.50
0.25
25µl
1.50
0.375
1.00
0.25
5µl
3.00
0.15
2.00
0.10
1020803000
50 - 300µl
5µl
300µl
0.70
2.10
0.25
0.75
150µl
1.00
1.50
0.50
0.75
50µl
1.50
0.75
0.80
0.40

Technical Specifications for Autoclavable Twelve Channel Pipette:

Catalog No.
Volume
Range
Dispensing
Increment

Test
Volume

Maximum
permissible
systematic
error
(inaccuracy)
Maximum
permissible
random error
(imprecision)
%
µl
%
µl
1021201000
0.5 - 10µl
0.1µl
10µl
1.50
0.15
1.50
0.15
5µl
2.50
0.125
2.50
0.125
1µl
4.00
0.04
4.00
0.04
1021202000
5 - 50µl
0.5µl
50µl
1.00
0.50
0.50
0.25
25µl
1.50
0.375
1.00
0.25
5µl
3.00
0.15
2.00
0.10
1021203000
50 - 300µl
5µl
300µl
0.70
2.10
0.25
0.75
150µl
1.00
1.50
0.50
0.75
50µl
1.50
0.75
0.80
0.40

 


                   
                   